Flammable Gas Control and Disposal Device Control the Gas Flow Rate

Product Name: Flammable Gas Control and Disposal Device Control the Gas Flow Rate

Product Model: GSL-KR200

Product Description: GSL-KR200 is a combustion gas module to dispose of residual flammable gas via a self-ignition heating coil. It is designed to pair with any SUPSEMI tube and atmosphere furnaces. It can also control the gas flow rate and communicate with a Honeywell hydrogen gas detector. It is a good choice for experiments under a reducing environment, such as hydrogen flow.

EMAIL

Product Description

GSL-KR200 is a combustion gas module to dispose of residual flammable gas via a self-ignition heating coil. It is designed to pair with any SUPSEMI tube and atmosphere furnaces. It can also control the gas flow rate and communicate with a Honeywell hydrogen gas detector. It is a good choice for experiments under a reducing environment, such as hydrogen flow.

Product Parameters

Power

• 200W

Voltage

• 110V, 208-240VAC, single phase, 50-60Hz

Features

• Applicable to dispose of various residual flammable gas along the exhaust line (designated gas needs to be discussed with us in advance).

• Compact, modularized, and compatible with various SUPSEMI furnaces. 

• Interlocks: in case of emergency, the gas flow and furnace heating will be cut off. 

   - a flammable gas leak is detected

   - Failure of a self-ignition heating coil

• One check valve to prevent gas backflow. 

• Two solenoid valve allows users to choose the flow type, inert or flammable.

Flow rate

• One float flow meter (100-1000 sccm, air)

Gas Inlets

• Two 1/4" compression tube fittings for both inert and flammable gas.


Hydrogen 

Detector 

System


• The hydrogen gas detector is connected externally. If alarmed, the hydrogen gas and furnace heating function will be cut off. 

• The alarm is preset at 20% LEL for hydrogen.

Furnace Dimension

360L × 340W × 400H (mm)

Gas Control and Disposal Device Dimension.jpg

Net Weight
5kg
Safety WARNING

• If equipped with a tube furnace, a metallic mesh cover must be installed at both ends of the furnace tube in case of emergency.

• The furnace and the combustion equipment are recommended to be placed in a well-ventilated place, or in a fume hood.

• The hydrogen detector must be installed.

• Operators must be trained by qualified technicians & follow the safety guideline of the laboratory.